Page #30 background imageLoading...
Page #30 background image
5730A
Operators Manual
1-16
Range Resolution
Frequency
(Hz)
Absolute / ±5 °C from calibration temperature Accuracy ±1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
±(ppm output + μV)
±(ppm output + mV)
2.2 V
1 μV
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
200 + 40
75 + 15
37 + 8
61 + 10
79 + 30
276 + 80
800 + 200
1300 + 300
220 + 40
80 + 15
39 + 8
63 + 10
81 + 30
300 + 80
900 + 200
1500 + 300
230 + 40
85 + 15
40 + 8
65 + 10
82 + 30
314 + 80
950 + 200
1600 + 300
240 + 40
90 + 15
42 + 8
67 + 10
85 + 30
336 + 80
1000 + 200
1700 + 300
200 + 40
75 + 15
25 + 8
55 + 10
80 + 30
230 + 80
700 + 200
1000 + 300
220 + 40
80 + 15
35 + 8
60 + 10
85 + 30
250 + 80
800 + 200
1100 + 300
22 V
10 μV
10 - 20
20 - 40
40 - 20k
20k - 50k
50k - 100k
100k - 300k
300k - 500k
500k - 1M
200 + 400
75 + 150
37 + 50
61 + 100
78 + 200
238 + 600
800 + 2000
1200 + 3200
220 + 400
80 + 150
39 + 50
63 + 100
80 + 200
243 + 600
900 + 2000
1300 + 3200
230 + 400
85 + 150
40 + 50
65 + 100
81 + 200
249 + 600
900 + 2000
1400 + 3200
240 + 400
90 + 150
42 + 50
67 + 100
83 + 200
254 + 600
1000 + 2000
1500 + 3200
200 + 400
75 + 150
25 + 50
55 + 100
80 + 200
250 + 600
700 + 2000
1100 + 3200
220 + 400
80 + 150
35 + 50
60 + 100
85 + 200
270 + 600
800 + 2000
1200 + 3200
220 V
[2]
100 μV
10 - 20
20 - 40
40 - 20 k
20 k - 50 k
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
200 +4
75 + 1.5
45 + 0.6
70 + 1
120 + 2.5
700 + 16
4000 + 40
6000 + 80
220 + 4
80 + 1.5
47 + 0.6
75 + 1
130 + 2.5
800 + 16
4200 + 40
7000 + 80
230 + 4
85 + 1.5
50 + 0.6
77 + 1
140 + 2.5
850 + 16
4300 + 40
7500 + 80
240 + 4
90 + 1.5
52 + 0.6
80 + 1
150 + 2.5
900 + 16
4400 + 40
8000 + 80
200 + 4
75 + 1.5
35 + 0.6
60 + 1
110 + 2.5
500 + 16
3600 + 40
6500 + 80
220 + 4
80 + 1.5
40 + 0.6
65 + 1
120 + 2.5
600 + 16
3800 + 40
7000 + 80
1100 V
[1]
1 mV
15 - 50
50 - 1 k
240 + 16
55 + 3.5
260 + 16
60 + 3.5
280 + 16
65 + 3.5
300 + 16
70 + 3.5
240 + 16
40 + 3.5
260 + 16
45 + 3.5
5725A Amplifier:
1100 V
1 mV
40 - 1 k
1 k - 20 k
20 k - 30 k
75 + 4
105 + 6
230 + 11
80 + 4
125 + 6
360 + 11
85 + 4
135 + 6
440 + 11
90 + 4
165 + 6
600 + 11
50 + 4
85 + 6
160 + 11
55 + 4
105 + 6
320 + 11
750 V
30 k - 50 k
50 k - 100 k
230 + 11
600 + 45
360 + 11
1300 + 45
440 + 11
1600 + 45
600 + 11
2300 + 45
160 + 11
380 + 45
320 + 11
1200 + 45
Notes:
1. Maximum output 250 V from 15-50 Hz.
2. See Volt-Hertz capability in Figure A.

Table of Contents

Other manuals for Fluke 5730A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5730A and is the answer not in the manual?

Fluke 5730A Specifications

General IconGeneral
TypeMultifunction Calibrator
Resistance Range0 to 1 GΩ
AC Voltage Frequency10 Hz to 1 MHz
DisplayColor LCD
InterfacesRS-232, Ethernet
Weight18 kg
DC Voltage Range0 to ±1100 V
Dimensions172 mm H x 432 mm W x 457 mm D (6.8 in H x 17 in W x 18 in D)