<bool> indicates a Boolean value (0 or 1) is required. The mnemonics OFF and ON
are also accepted for 0 and 1, respectively.
<conv> indicates a conversion mnemonic is required.
<param> indicates a parameter name is required.
<seri> indicates a serial number is required.
<res> indicates a resistance value is required.
<volt> indicates a voltage value is required.
<unit> indicates a temperature unit is required.
<temp> indicates a temperature °C/F is required.
<pass> indicates a password is required.
<port> indicates a port number is required.
<label> indicates an eight character label is required.
<year> indicates a four digit number is required.
<month> indicates a one or two digit number is required.
<day> indicates a one or two digit number is required.
<hour> indicates a one or two digit number is required.
<minute> indicates a one or two digit number is required.
<second> indicates a one or two digit number is required.
<baud> indicates a valid baud number is required.
*CLS
Clear the status registers.
Example: *CLS
Response: None
*IDN?
Gets manufacturer, model number, serial number and firmware version of the product.
Example: *IDN?
Response:
HART,9190A,A79002,1.00
*OPT?
Read the product configuration, reference hardware enabled (1) or not (0) (see
SYST:CONF:MOD). This command is a read only command and returns the state of the
reference functionality (0, 1).
Example: *OPT?
Response: 1