Remote Operation
Remote Operation Commands 3
3-15
CAL:DATE:MOD[?] [<year>,<month>,<day>] (-P model only)
Note
This command is unconditionally protected, that requires a password to set it.
Read or set the calibration date for the -P Module (INPUT CALibration DATE). The entered
values are all numeric and “yyyy” is a four digit year (2000-2135), “mm” is a two digit month
(1-12), and “dd” is a two digit day (1-31).
Read Example: CAL:DATE:MOD?
Response: 2007,05,24
Set Example: CAL:DATE:MOD 2007,12,30
CAL:DATE:UNIT[?] [<year>,<month>,<day>]
Note
This command is unconditionally protected, that requires a password to set it.
Read or set the calibration date for the heat portion of the Calibrator. The entered values are all
numeric and “yyyy” is a four digit year (2000-2135); default 2007, “mm” is a two digit month
(1-12), and “dd” is a two digit day (1-31).
Read Example: CAL:DAT:UNIT?
Response: 2007,05,24
Set Example: CAL:DAT:CAL 2006,12,30
CALC1:CONV:CAT? (-P model only)
Read the list of reference probe characterization methods: “CVD, I90, IEC, RES”. Supplies the
list of PRT/RTD characterization methods available.
Example: CALC1:CONV:CAT?
Response: “CVD”,”I90”,”IEC”,”RES”
CALC1:CONV:DATE[?] [<yyyy,mm,dd>] (-P model only)
Note
This command is unconditionally protected, that requires a password to set it.
Read or set the INFO-CON probe calibration date in yyyy,mm,dd format.
Year Range = {2000 – 2135}; Default: 2007
Month Range = {1 – 12}; Default: 1
Day Range = {1 – 31}; Default: 1
Read Example: CALC1:CONV:DATE?
Response: 2007,10,09
Set Example: CALC1:CONV:DATE 2007,09,06
CALC1:CONV:NAME[?][n] (-P model only)
Note
This command is unconditionally protected, that requires a password to set it.
Read or set the reference probe characterization method. “n” is an alpha numeric value the
represents: CVD, I90, IEC, or RES.
Range={ITS90, CVD, IEC, RES} or 0-3, Default:ITS90.
Read Example: CALC1:CONV:NAME?
Response: CVD
Set Example: CALC1:CONV:NAME I90