Chapter 2 Working with DaqLab 63
e. If not selected, click the Set Offset tab.
f. Ensure that All is selected in the Input drop-down menu.
g. Enter the offset value in the Offset text box (if the two offset values are not
identical enter the average of the two values).
h. Click OK.
Setting the offset to a specific input
After setting the offset to all inputs simultaneously, you can set the offset of each input separately for
even better accuracy.
To set the offset of a specific input:
a. Measure two known reference values as described in Measure two reference
values above, making sure to use the input you need to calibrate.
b. Enter the offset value as described in Enter the offset value above, making sure
to select the correct input number in step f, instead of All.
c. You can repeat the procedure for every input.
Zero Calibration
a. Connect DaqPRO to the PC.
b. Click Logger on the menu bar, then click Calibration > Calibrate Sensors.
c. Enter the calibration password, then click OK.
d. Choose a sensor from the Select sensor drop-down menu and click OK to open
the Sensor calibration dialog box: