13
Click [Next >]
The preliminary settings / general measurement data are displayed:
-
Ambient temperature [° C]
-
Mains voltage [V]
-
Open circuit voltage U
S
[V]
If no data is displayed, it can be found by clicking on [Start measuring pro-
cess].
If there is still no data displayed, check the F2 and F3 fuses on the back of
the calibration system.
14
Carry out a visual inspection of the welding system and the calibration sys-
tem, according to the calibration instructions
15
If the visual inspection does not reveal any defects, select [Click here]
16
Click [Next >]
The summary of the calibration process is displayed.
17
Define the calibration class:
o Precision
o Standard
18
Click [Start automatic calibration process]
The calibration process starts.
When a calibration process is active
-
A corresponding notification is displayed
-
The [Interrupt calibration] button is available
-
All measuring points are checked one after the other
-
Current measured values are displayed
-
A progress bar with the remaining time is displayed
If the calibration process is interrupted before the welding wire has been cal-
ibrated, a corresponding notification is displayed.
19
Change over the welding system for calibration of the welding wire:
(see page 45 onwards)
-
Connect welding wire calibration system and calibration system
-
(+) Disconnect power cable from welding system
-
Connect the welding torch to the welding system
-
Remove the gas nozzle from the welding torch
-
Thread the wire
-
Thread the welding wire into the sensor
-
Secure the welding torch in the welding wire calibration system
57