Set the toggle switch to A (current) Set the selector switch on the interior of the
device to 2-step mode.
19
Click [Start measuring process]
20
Carry out a visual inspection of the welding system and the calibration sys-
tem, according to the calibration instructions
21
If the visual inspection does not reveal any defects, select [Click here]
22
Click [Next >]
The summary of the calibration process is displayed.
23
Define the calibration class:
o Precision
o Standard
NOTE!
In the case of step-switched machines, the set values can be adapted manually
in the calibration program.
24
Set the stage switch on the power source as specified in the calibration pro-
gram for measuring point M1 (e.g. stage A1)
25
Click [Start calibration process]
26
Read the measured value for the respective level in the calibration program
and enter it in the Set value [A] box
27
Click [Start calibration process]
28
Enter the value displayed on the power source in the Display [A] box
29
Click [Next >]
30
Repeat steps 24 - 29 for measuring points M2 - M5
For the subsequent measurement of M5 to M1, the set values entered are
saved by the calibration program.