6-16 250 Series Maternal/Fetal Monitor Revision D
2020551-001
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
20. Pull red tubes through chassis grommet away from Front-bezel.
21. Disconnect orange connector receptacle of power switch cable assembly from
orange chassis-mounted plug.
22. Remove front bezel from chassis tabs.
23. Remove Ultrasound board from Front-end Motherboard. Remove SpO
2
Carrier
board assembly from Front-end Motherboard.
24. Remove MECG/FECG board assembly from Front-end Motherboard.
25. Remove Isolated Power Supply board from Front-end Motherboard.
26. Remove three insulating metal sheets from Front-end Motherboard.
27. Remove five screws fastening the Front-end Motherboard to the chassis.
28. Remove Front-end Motherboard.
29. Disconnect 20-conductor Recorder ribbon cable from Main board connector J9.
30. Disconnect power supply cable from J1 connector on Recorder board.
31. Disconnect Recorder EMC ground wire Fast-on tab from chassis.
32. Remove four screws fastening Recorder assembly mounting bracket to chassis.
33. Remove Recorder assembly and mounting bracket.
34. Remove two screws fastening Pneumatics assembly to chassis.
35. Disconnect clear tubing from both PT1 and PT2 on the Main board.
36. Remove Pneumatics assembly from the Main board.
37. Remove four screws fastening the Comm. Board to the chassis.
38. Remove Comm. board from Main board.
39. Disconnect power supply cable from Main board connector J8.
40. Disconnect speaker cable from Main board connector J10.
41. Remove six screws that fasten Main board to chassis.
42. Remove Main board with EMI plate from chassis.
43. Cut tie-wrap holding power-switch cable to cable-tie mount on chassis.
44. Remove power supply chassis-mounted connector from chassis.
45. Remove cable-tie mount from side of chassis.
46. Disconnect Power Supply cable from J1 connector on Recorder board.
47. Disconnect Power Supply cable from J8 connector on Main board.
48. Remove three flat-head screws from side of Power Supply.
49. Remove four pan-head screws from back of Power Supply.
50. Remove one pan-head screw from inside bottom of Power Supply.
51. Remove Power Supply from chassis.
52. Replace chassis assembly with new chassis assembly.
53. Place Power Supply onto chassis.
54. Fasten three flat-head screws into side of Power Supply.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250 Series and is the answer not in the manual?

GE Corometrics 250 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 250 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals