Revision D 250 Series Maternal/Fetal Monitor 6-17
2020551-001
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
55. Fasten four pan-head screws into back of Power Supply.
56. Fasten one pan-head screw into inside bottom of Power Supply.
57. Insert power supply power-switch connector into chassis.
58. Adhere new cable tie mount to side of chassis.
59. Use tie-wrap to fasten power-switch cable to cable-tie mount on chassis.
60. Place Main board with EMI plate onto chassis. Use six screws to fasten Main
board to chassis.
61. Connect speaker cable to Main board connector J10.
62. Connect power supply cable to Main board connector J8.
63. Place Comm. board onto Main board.
64. Fasten Comm. board onto chassis with four screws.
65. Place new Pneumatics assembly onto Main board
66. Insert two screws fastening Pneumatics assembly onto chassis.
67. Connect clear section of tubing from E1 to PT1 on Main board.
68. Connect clear section of tubing from E2 to PT2 on Main board.
69. Pull red tubes through chassis grommet towards Front-bezel.
70. Fasten Recorder assembly and mounting bracket to chassis with four M3.5x6
screws.
71. Connect Recorder EMC ground wire Fast-on tab to chassis.
72. Connect power supply cable to J1 connector on Recorder board.
73. Connect 20-conductor Recorder ribbon cable to Main board connector J9.
74. Slide new Front-end Motherboard under card cage onto chassis.
75. Fasten to chassis with five screws.
76. Insert three insulating metal sheets onto Front-end Motherboard.
77. Insert Isolated Power Supply board onto Front-end Motherboard.
78. Insert MECG/FECG board assembly onto Front-end Motherboard.
79. Insert SpO
2
Carrier board assembly onto Front-end Motherboard.
80. Insert Ultrasound board onto Front-end Motherboard.
81. Insert front bezel into chassis tabs.
82. Connect orange connector receptacle of new power switch cable to orange
chassis-mounted plug.
83. Pull red tubes through chassis grommet towards Front-bezel.
84. Connect red tubes to Front-bezel NIBP connector.
85. Connect IUP cable to Front-end Motherboard.
86. Connect FECG cable to Front-end Motherboard.
87. Place FECG ground cable onto chassis post and tighten with nut.
88. Insert MECG and FECG cable beads into chassis bead clips.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250 Series and is the answer not in the manual?

GE Corometrics 250 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 250 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals