Preventive maintenance: Cleaning, disinfecting and storing GE ECG cables and leadwires
5-8 Dash 3000/4000/5000 2000966-542D
Sterilization
NOTE
EtO sterilization is not recommended, but may be required for cables 
and leadwires. Frequent sterilization will reduce the useful life of 
cables and leadwires.
Sterilize with ethylene oxide gas (EtO) at a maximum temperature of 50° 
C/122° F. After EtO sterilization, follow the recommendations from the 
sterilizer manufacturer for required aeration.
Cautions
 Never immerse the handheld device, cables, or leadwires in any 
liquid.
 Do not pour or spray any liquid directly on cables or leadwires or 
permit fluid to seep into connections or openings.
 Never use conductive solutions, solutions that contain chlorides, wax, 
or wax compounds to clean handheld devices, cables or leadwires.
 Never use solutions or products that contain the following:
 Any type of Ammonium Chloride such as, but not limited to: 
Dimethyl Benzyl Ammonium Chloride or Quaternary 
Ammonium Chloride solutions
 Abrasive cleaners or solvents of any kind
 Acetone
 Ketone
 Betadine
 Alcohol-based cleaning agents
 Sodium salts
 Never autoclave or steam clean cables or leadwires.
Storage
 Store in a dry well-ventilated area.
 Vertically hang cables and leadwires.
 Do not coil leadwires or cables tightly around any medical device.
Improper cleaning products and processes impact or results
 Product discoloration
 Metal part corrosion
 Brittle wires
 Brittle and breaking connectors
 Reduced cables and leadwires life
 Unit malfunction
 Void warranty