Functional and electrical safety checks: Functional Checkout procedures
8-20 Dash 3000/4000/5000 2000966-542D
Functional Checkout procedures
Frequency
Qualified personnel must perform the checkout procedures:
 Every 12 months after receipt of the device (Preventative 
Maintenance). Refer to “Maintenance schedule” on page 5-2 for more 
information.
 Each time the main enclosure is disassembled or a circuit board is 
removed, tested, or replaced (Corrective Maintenance). Refer to 
“Recommended checkout” on page 7-52 for more information.
Identify enabled patient parameters and software options
The check out procedures support all enabled patient parameter and 
software options. To view the options enabled on the patient monitor, go 
to the Main Menu and select MORE MENUS > MONITOR SETUP > 
SOFTWARE CONFIGURATION.
The following is a list of all available patient parameter and software 
options.
 0BP (invasive)
 1BP or 2BP (invasive)
 3BP or 4BP (invasive)
 12SL
 ACI-TIPI
 HI-RES TRENDS
 ETCO
2
 (CO
2
) (End-tidal CO
2
)
 NETWORK 
 AVOA Plus
 CARDIO-PULMONARY
 CARDIAC