Configuration: Service Mode settings
4-12 Dash 3000/4000/5000 2000966-542D
Patient-Monitor Type
The Patient-Monitor Type selection determines the type of patient 
monitor desired, e.g., adult, neonatal or operating room. Different alarms 
and parameters are activated for each selection.
CAUTION
Each time the patient-monitor type is changed, the 
ADMIT MENU function defaults to STANDARD 
configuration. Be aware that some patient alarms and 
parameters may be changed.
NOTE
The keypad/remote control is DIDCA programmed for specific 
patient monitor types. The error message, “WARNING: REMOTE 
MISMATCHED WITH MONITORING MODE” displays if the 
patient monitor and keypad/remote control do not match.
1. Enter service mode (See “Service Mode settings” on page 4-11) and 
select PATIENT-MONITOR TYPE. Be sure to read the information 
in the Attention box before changing anything.
2. Rotate Trim Knob control to select the type of environment the 
patient monitor will be used in. 
3. Press Trim Knob control to exit. Your selection displays at the top 
left of the screen after the date.