Electromagnetic compatibility (EMC): Electromagnetic compatibility (EMC)
A-6 Dash 3000/4000/5000 2000966-542D
Recommended separation distances
The table below provides the recommended separation distances (in 
meters) between portable and mobile RF communications equipment and 
the Dash 3000/4000/5000.
The Dash 3000/4000/5000 is intended for use in the electromagnetic 
environment on which radiated RF disturbances are controlled. The 
customer or the user of the Dash 3000/4000/5000 can help prevent 
electromagnetic interference by maintaining a minimum distance 
between portable and mobile RF communications equipment 
(transmitters) and the Dash 3000/4000/5000 as recommended below, 
according to the maximum output power of the communications 
equipment.
For transmitters rated at a maximum output power not listed above, the 
recommended separation distance [d] in meters (m) can be estimated 
using the equitation applicable to the frequency of the transmitter, 
where P is the maximum output power rating of the transmitter in watts 
(W) according to the transmitter manufacturer.
NOTE:
These guidelines may not apply in all instances. Electromagnetic 
propagation is affected by absorption and reflection from structures, 
objects and people.
Separation distance estimation
(Transmit power of 802.11a/b/g wireless communication device) 
Power of the Source 
(Watts)
Separation Distance (Feet) Separation Distance 
(Meter)
0.01 0.75 0.23
0.1 2.39 0.73
17.542.3
10 23.94 7.3
100 75.44 23