Functional and electrical safety checks: Electrical safety tests
2000966-542D Dash 3000/4000/5000 8-5
Power outlet test
Verify that the power outlet is wired correctly per the country’s electrical 
code standard before starting the following electrical safety tests. The 
results of the following tests will be inaccurate unless a properly wired 
power outlet is used. Use only non-isolated power outlets when 
performing safety tests.
Power cord and plug
Verify the power cord being used with the patient monitor is good. The 
following are a couple of things to check for in this regard:
 Failure of the power cord strain relief is very common. Often times 
users of the equipment pull on the power cord itself, rather than the 
power cord plug, to unplug the patient monitor from a wall 
receptacle. If in doubt, test for continuity through each conductor of 
the power cord connector and plug.
 Verify line, neutral, and earth conductors are properly connected to 
the power cord plug and are not short-circuited. Replace the power 
cord, as necessary with a regulatory-approved cord for the country of 
use. 
WARNING
Use only AC power cords recommended or manufactured 
by GE.