Configuration: Advanced user procedures
4-24 Dash 3000/4000/5000 2000966-542D
Transfer monitor defaults
NOTE
Transfer Monitor Defaults feature is part of the CARESCAPE 
Network software option and is available in software version 3 or 
later. You must purchase and enable this software option before you 
can transfer patient monitor defaults. 
NOTE
To prevent the accidental erasure of the client patient monitor’s 
wireless LAN defaults, make sure that the server and client patient 
monitors’ wireless LAN defaults match before transferring patient 
monitor defaults.
The patient monitor defaults set on one patient monitor (server) may be 
transferred to another patient monitor (client), provided that the client 
patient monitor shares the following: 
 the same software revision (version), 
 the same patient-monitor type (e.g., ADULT, NEO, or OR), 
 the same wireless LAN configuration,
 the same software package (e.g., Basic, Cardiac, Cardiopulmonary), 
and 
 the same country code (e.g., Default or France).
The following defaults transfer when using this feature:
 all patient monitor defaults
 wireless LAN configuration
 custom default names
 12 SL location
 12 SL site