Network disclosure to facilitate network risk management: 
C-6 Dash 3000/4000/5000 2000966-542D
Potential risks to safety, effectiveness or security resulting from 
failure of IT network to provide the required characteristics
WARNING
Connection of the Dash patient monitor to a network that 
includes other equipment could result in other 
unidentified risks to patients, operators or third parties. 
The responsible organization should identify, analyze, 
evaluate and control these risks on an ongoing basis 
including after changes to the network such as those 
listed below, which could introduce new risks and require 
additional analysis.
 Changes in network configuration.
 Connection of additional items to the network.
 Disconnecting items from the network.
 Update of equipment connected to the network.
 Upgrade of equipment connected to the network.