Installation: Connections
2000966-542D Dash 3000/4000/5000 3-7
Defib sync
The connector provides ECG analog output signals to user-supplied 
equipment.
CAUTION
Equipment damage. Connect all peripheral equipment 
before plugging the power cord into an AC outlet. 
Otherwise, connectors may be damaged.
AC power
Use this connector to apply power to the patient monitor. The patient 
monitor will be powered at all times when using AC power (there is no 
AC power switch). The patient monitor is preset at the factory for a 
specific line voltage. Before applying power, verify the patient monitor’s 
line voltage setting matches your power supply. If required, use a small 
bladed screw driver to change the line voltage setting. Refer to the label 
on the back of the unit for the voltage and current requirements.
CAUTION
Verify the line voltage selector setting before plugging 
the power cord into an AC outlet. The power supply may 
be damaged if the power source applied is different from 
the line voltage setting.
Power up
After making all connections, plug the power cord into an AC wall outlet.
When all cables are properly connected, press the Power button to turn 
the patient monitor on. All front panel indicators will illuminate until 
the power-up sequence is complete. After approximately 20 seconds you 
should see a display on the screen.
Configure
Go to “Configuration” on page 4-1 to configure and set up the Dash 
patient monitor for patient use.
115V
line voltage selector:
115V for a 100-120V power source
230V for a 220-240V power source