Troubleshooting: Battery alarms and messages
2000966-542D Dash 3000/4000/5000 6-7
Battery alarms and messages
Battery alarms occur when one of the following battery conditions occur: 
 Low battery
 Empty battery 
 Battery failures
 Charger failures
These Battery messages help identify the cause of a battery alarm. 
Battery Messages appear in three different locations of the patient 
monitor:
 ECG waveform area.
 Battery Status Information window.
 Battery Fuel Gauge icon.
Battery messages displayed in the ECG waveform area
The following battery messages are displayed in the ECG Waveform 
Area.
Battery messages displayed in the ECG waveform area
Message Battery alarm Cause
BATTERY LOW  System WARNING 
Critical Low Battery
Only 10 minutes per battery of run 
time remaining (10 minutes if one battery, 20 minutes if two 
batteries).
POWERING DOWN System WARNING
Empty Battery
There is no battery run time remaining.
CHECK BATT STATUS System MESSAGE
Battery Failure
A minor failure has occurred while using or 
charging the battery.
BATTERY ERROR System WARNING
Battery Failure
A serious failure has occurred while using 
or charging the battery. An incompatible battery may be 
installed.
CHECK BATT STATUS
NOTE
INTERNAL CHARGER 
FAILED, CALL SERVICE 
also appears in the Battery 
Status information window.
System MESSAGE
Charger Failure
Charger communications have failed.