Field replaceable units: Replacement procedures
2000966-542D Dash 3000/4000/5000 7-9
Replacement procedures
Unless otherwise stated, reassemble the patient monitor in reverse order 
of disassembly.
NOTE
When reassembling the patient monitor, observe the following 
guidelines:
 To prevent cracking the plastic components, do not overtighten 
the screws fastening the plastic components together.
 To prevent cross-threading plastic components, carefully start 
threading the screws by hand, then finish tightening the screws 
using an appropriate hand tool.