Functional Checks
C
O
N
F
I
N
S
T
C
M
P
M
P
A
S
S
F
A
I
L
Monitor Power Up - (POST)
  
ECG 5 LD (S)
ECG 12SL 
(option)
Respiration (S)
Temperature 1 and 2  (S)
Cardiac Output 
(option)
Invasive BP 
(option)
Pulse Oximetry (S)
NBP (S)
Analog Out, Defibrillator 
Synchronization
(S)
End-Tidal CO2 
(option)
Batteries
Battery A 
(option)
Battery B
(option)
Printer
(option)
Display (S)
Speaker  (S)
Network 
(option)
    
Remote Control 
(option)
BISx
(option)
Wireless LAN 
(option)
    
Dash Port 2 Docking 
Station 
(option)
TRAM-RAC 2A w/ICG 
(option)
  
TRAM-RAC 2A w/SAM 
(option)
  
Abbreviations 
DUT  Device Under Test 
Standard (S)  Standard configuration test required 
Option (O)  Test as required if the option is installed 
INST Installation 
CM Corrective Maintenance 
PM Preventive Maintenance 
CONF  Configuration – Standard (S) or Option (O) 
Device Test Results  Tested By  Date 
Pass (9) Fail (9)
   
Step
DUT
Done 
9
Electrical Safety Tests 
Test 
Results
Expected 
Limits
I
N
S
T
C
M
P
M
P
A
S
S
F
A
I
L
A Ground Continuity  Ohms  Ohms 
1.  Measured Resistance  
 <.2     
B Earth Leakage Current 
µ
A
µ
A
USA
1.  Pol Norm, Neut Clsd   
300 500 
   
2.  Pol Norm, Neut Open   
300 1000 
   
3.
Pol Rev, Neut Open   
300 1000 
   
4.
Pol Rev, Neut Clsd   
300 500 
   
C Enclosure Leakage Current 
1.  Pol Norm, Neut Clsd, Gnd Clsd   
100
   
2.  Pol Norm, Neut Open, Gnd Clsd   
300
   
3.  Pol Norm, Neut Clsd, Gnd Open 
 300     
4.  Pol Rev, Neut Clsd, Gnd Open 
 300     
5.  Pol Rev, Neut Open, Gnd Clsd 
300
   
6.  Pol Rev, Neut Clsd, Gnd Clsd 
100
   
D
Patient Leakage Current
to Ground
ECG 
SP
O2
µ
A
1.  Pol Norm, Neut Clsd, Gnd Clsd 
  10      
2.  Pol Norm, Neut Open, Gnd Clsd 
  50      
3.  Pol Norm, Neut Clsd, Gnd Open 
  50      
4.  Pol Rev, Neut Clsd, Gnd Open 
50
   
5.  Pol Rev, Neut Open, Gnd Clsd 
50
   
6.  Pol Rev, Neut Clsd, Gnd Clsd  10        
E.
Patient Leakage Current 
Mains on Applied Part 
ECG 
SP
O2
µ
A
1.  Pol Norm, Neut Clsd, Gnd Clsd
  50      
2.  Pol Rev, Neut Clsd, Gnd Clsd 
  50      
F.
BISx Patient Leakage Current
to Ground (option)
µ
A
µ
A
1.  Pol Norm, Neut Clsd, Gnd Clsd 
100
   
2.  Pol Norm, Neut Open, Gnd Clsd 
500
   
3.  Pol Norm, Neut Clsd, Gnd Open 
500
   
4.  Pol Rev, Neut Clsd, Gnd Open 
  500 
   
5.  Pol Rev, Neut Open, Gnd Clsd 
  500 
   
6.  Pol Rev, Neut Clsd, Gnd Clsd 
  100 
   
G.
BISx Patient Leakage Current 
Mains on Applied Part (option) 
µ
A
µ
A
1.  Pol Norm, Neut Clsd, Gnd Clsd
    5000    
2.  Pol Rev, Neut Clsd, Gnd Clsd 
    5000