Troubleshooting: Monitor defaults transfer
2000966-542D Dash 3000/4000/5000 6-11
Monitor defaults transfer
Below is a list of error messages that may display when performing a 
patient monitor defaults transfer.
Storing monitor defaults
Copying stored monitor defaults
Message Description
ERROR - UNIT DEFAULTS HAVE NOT 
BEEN SAVED
Patient monitor defaults were not saved on 
the server patient monitor due to a memory 
error in the patient monitor.
Message Description
ERROR COPYING UNIT DEFAULTS - 
NETWORK ERROR
Either an error has occurred while the 
defaults were being transferred, or defaults 
have not been saved on the server patient 
monitor.
ERROR COPYING UNIT DEFAULTS - 
UNIT TYPE MISMATCH
The server and client patient monitors are 
set to different unit types (e.g., ADULT, 
NEO, or OR)
ERROR COPYING UNIT DEFAULTS - 
SOFTWARE VERSION MISMATCH
The server and client patient monitors have 
different software versions (e.g., V2C, V3A).
ERROR COPYING UNIT DEFAULTS - 
DATA CORRUPTION ERROR
The patient monitor defaults transferred have 
become corrupted, possibly due to a 
memory error.
ERROR COPYING UNIT DEFAULTS - 
SOFTWARE FEATURE LEVEL 
MISMATCH
The server and client patient monitors are 
set to different software levels (e.g., Basic, 
Cardiac, Cardiopulmonary).
ERROR COPYING UNIT DEFAULTS - 
COUNTRY CODE MISMATCH
The server and client patient monitors are 
set to different country codes (e.g., 
DEFAULT or FRANCE).