Equipment overview: Components
2-10 Dash 3000/4000/5000 2000966-542D
Verify compatibility of an unmarked battery as follows.
1. Install a battery pack in the patient monitor.
2. Using the Trim Knob control, access the Service Mode menu 
starting from the Main Menu. Select MORE MENUS > MONITOR 
SETUP > SERVICE MODE.
3. Enter password using the Trim Knob control to select the day and 
month from patient monitor screen with leading zeros. (e.g. July 4 = 
0407).
4. Select BATTERY SERVICE.
5. Verify that the MANUFACTURER NAME does not display 
INCOMPAT, NME, or UNKNOWN for the battery corresponding to 
BATTERY A or BATTERY B slot.
Optional components
TRAM-rac 2A module housing
The TRAM-rac 2A module housing currently supports the SAM and ICG 
modules.
An integral power supply is used to run the TRAM-rac 2A and support 
the needed voltages.
Dash Port 2 docking station
The docking station is a quick mount/dismount base for a Dash patient 
monitor. It gives the patient monitor easy connect or disconnect access to 
AC power, CARESCAPE Network, a remote display, and auxiliary 
devices.
See the Dash Port 2 Docking Station Operating Instructions and the 
Dash Port 2 Docking Station Service Manual for additional information.
NOTE
When a Dash patient monitor is connected to the docking station, 
only the docking station’s Ethernet port is active. The Dash patient 
monitor’s network port remains inactive until the patient monitor is 
disconnected from the docking station.