Configuration: Boot Code settings
2000966-542D Dash 3000/4000/5000 4-19
Set CIC and QS protocol
The CIC and QS Protocol default setting is Seg50/51 (Segment 50/51) 
and should not be changed. However, future products may require 
Hilltop protocol. Use the Boot Code Service Menu to change the CIC 
Protocol and QS Protocol.
Set MUSE system protocol
The patient monitor transmits 12SL and ACI-TIPI data over the optional 
CARESCAPE Network to the MUSE System and ST Guard. The formats 
used for this process are Hilltop and Segment 50/51. Since the Segment 
50/51 format does not support 500 Hz ECG data or ACI-TIPI, records 
originally stored on the MUSE System in Hilltop format cannot display 
at the ST Guard. The default setting is Hilltop.
1. Activate Boot Code (“Boot Code settings” on page 4-18) and select 
SET CONFIGURATION menu option.
2. Select MUSE System Protocol then choose Hilltop or Seg50/51.
 Hilltop—Sends 12SL records to the MUSE System in Hilltop 
format. The 12SL records contain 500 samples per second of ECG 
data. ST Guard will not be able to retrieve 12SL records stored on 
the MUSE System in Hilltop format.
 Seg50/51—Sends 12SL records to the MUSE System in Segment 
50/51 format with MAC Rhythm statements. The 12SL records 
stored on the MUSE System contain 240 samples per second of 
ECG data. ST Guard can retrieve 12SL records from the MUSE 
System. ACI-TIPI data can not be retrieved at the MUSE System 
if this protocol is selected.
3. Select Exit and wait for the patient monitor to display the message, 
“Writing settings to Flash config mem...”
4. Manually reboot the patient monitor to implement this change.
NOTE
Before you can transmit data to the MUSE System and ST 
Guard, you must set up the MUSE system’s Location ID and Site 
Number.
Do not configure MUSE system’s location at Site 0. Site 0 
appears invalid to MUSE and any message coming from a device 
on Site 0 is dumped.
5. Refer to “Service Mode settings” on page 4-11 and select MONITOR 
SETTINGS > SET LOCATION, then select SET SITE NUMBER.
Transcutaneous Pace Blank Length
This menu option is reserved for future use and should not be changed.