Functional and electrical safety checks: BISx (option) current leakage tests
2000966-542D Dash 3000/4000/5000 8-17
3. Connect the BISx power supply cord to the power outlet on the
leakage tester.
4. Turn the BISx ON by connecting the AC power adaptor.
5. Read the leakage current indicated on the tester.
6. Change the leakage tester switches to:
Polarity – NORMAL
Neutral – OPEN
GND (Earth) – CLOSED
7. Read and record the leakage current indicated on the tester.
8. Change leakage tester switches to:
Polarity – NORMAL
Neutral – CLOSED
GND (Earth) – OPEN
9. Read and record the current leakage indicated on the tester.
10. Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – OPEN
11. Read and record the current leakage indicated on the tester.
12. Change leakage tester switches to:
Polarity – REVERSED
Neutral – OPEN
GND (Earth) – CLOSED
13. Read and record the current leakage indicated on the tester.
14. Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – CLOSED
15. Read and record the current leakage indicated on the tester.
16. Set the power switch of the device to OFF.
17. Leave the test configuration set up for the sink leakage test.
Acceptance criteria NC
If any reading with the ground and neutral closed is greater than 100 uA,
the BISx fails this test. Contact GE Technical Support.