Equipment overview: Theory of operation
2-26 Dash 3000/4000/5000 2000966-542D
The DAS block diagram with Nellcor 05 DAS PCB assembly consisting of 
three sections:
 
The DAS design includes the Nellcor MP100 Module technology. Nellcor 
SPO2 is classified as motion-resistant. It uses advanced digital signal 
processing algorithms to extract very low-level SPO2 signals in the 
presence of artifact induced noise.
The DAS contains all circuitry necessary to support the MP100 Module. 
The Nellcor MP100 Module contains: 
 All SPO2 front end circuitry, such as the digitally programmable 
photodetector signal conditioning, A/D conversion, digitally 
programmable back-to-back LED IR/RED emitter drive and on-board 
diagnostic / sensor identification signal processing circuits. 
 A digital signal processor (DSP), clock circuitry, and program 
memory. The Nellcor SPO2 algorithms executes on the DSP. No 
external memory interface is provided. The DSP is "reset-able" by the 
DAS CPU. 
 A UART for communication of commands and data.
Parameter functions of the DAS with Nellcor OxiMax SPO2, 
four BP channels and Dinamap SuperSTAT NIBP
Except for the SPO2 parameter and additional 2 IBPs, the DAS with the 
Nellcor OxiMax SPO2 parameter supports the same parameters as DAS 
with the GE Ohmeda SPO2 parameter. Refer to “ECG” on page 2-20.
Probes and cable
The DAS is intended for use only with Nellcor digital probes. It is not for 
use with non-Nellcor probes or with Nellcor R-cal (resistor id) probes. An 
adapter cable is used to interface the probes to the DAS front panel 
Nicolay connector. 
Section Description
ECG/Respiration Separately isolated section includes a 7.3728MHz 68HSC05 
microcontroller, A/D conversion, signal processing hybrids and 
DC-DC isolation converter.
Main DAS Separately isolated section includes a 22.1184MHz Motorola 
MC68332 microcontroller with FLASH and SRAM memory, an 
Analog Devices ADSP-21062 32-bit floating point digital signal 
processor running at 12.096MHz, A/D conversion, signal 
processing hybrids/modules and DC-DC isolation converter.
Non-isolated circuits Includes serial and parallel host interfaces and NBP pump, valves, 
and over-pressure circuitry.