General Procedures and Functional Checks
4-40 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
CWD Functional Check (continued)
Activating CW Doppler
To activate CW Doppler Mode:
1. Ensure that the appropriate CW probe is connected.
2. Set the hot key (From programmable key F5-F12) for CWD.
3. Press the hot key which is assigned for CWD mode. The
Doppler Spectrum appears, along with the CW Touch Panel
Menu.
The following CW parameters are displayed: Frequency, Gain,
Acoustic Output, Scale, Wall Filter and Dynamic Range.
Exiting CW Doppler
To exit CW Doppler Mode:
Press the hot key which is assigned for CWD Mode.