Functional checks
LOGIQ e – Basic Service Manual 4-41
5461614-100 English Rev. 6
Basic Measurements
NOTE: The following instructions assume that you first scan the patient
and then press Freeze.
Distance and Tissue Depth Measurements
1. Press Measure once; an active caliper displays.
2. To position the active caliper at the start point (distance) or
the most anterior point (tissue depth), move the Trackball.
3. To fix the start point, press Enter. The system fixes the first
caliper and displays a second active caliper.
4. To position the second active caliper at the end point
(distance) or the most posterior point (tissue depth), move
the Trackball.
5. To complete the measurement, press Enter. The system
displays the distance or tissue depth value in the
measurement results window.
Circumference/Area (Ellipse) Measurement
1. Press Measure once; an active caliper displays.
2. To position the active caliper, move the Trackball.
3. To fix the start point, press Enter. The system fixes the first
caliper and displays a second active caliper.
4. To position the second caliper, move the Trackball.
5. Adjust the Ellipse; an ellipse with an initial circle shape
appears
6. To position the ellipse and to size the measured axes (move
the calipers), move the Trackball.
7. To increase the size, rotate the Ellipse button clockwise. To
decrease the size, contrarotate the Ellipse button
8. To complete the measurement, press Enter. The system
displays the circumference and area in the measurement
results window.
Worksheets
Measurement/Calculation worksheets are available to display
and edit measurements and calculations. There are generic
worksheets as well as Application specific worksheets. The
worksheets are selected from the Measurement Touch Panel.