General Procedures and Functional Checks
4-42 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Probe/Connectors Check
NOTE: Probes can be connected at any time, whether the unit is ON or
OFF
Connecting a Probe
1. Place the probe's carrying case on a stable surface and
open the case.
2. Carefully remove the probe and unwrap the probe cable.
3. DO NOT allow the probe head to hang free. Impact to the
probe head could result in irreparable damage.
4. Align the connector with the probe port and carefully push
into place.
5. Lock the probe latch upward.
6. Carefully position the probe cord so it is free to move and is
not resting on the floor.
Activating the probe
The probe activates in the currently-selected operating mode.
The probe's default settings for the mode and selected exam are
used automatically.
Deactivating the probe
1. Press the Freeze key.
2. Gently wipe the excess gel from the face of the probe.
(Refer to the Basic User Manual for complete probe
cleaning instructions).
3. Carefully slide the probe around the right side of the
keyboard, toward the probe holder. Ensure that the probe is
placed gently in the probe holder.