Replacement Procedures
8-22 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Reload the Correct Preset Region
NOTE: After the system software loading completion, please reload the
correct preset region.
1. Reboot the system.
2. Select Utility -> General.
3. In the Location portion, select the right region in the field of
Preset Region.
Figure 8-10. Preset Region
4. Reboot the system.