10-14 MAC 1200/1200 ST Revision C
2012250-021
System Setup: Stress Test Mode
Push to open the Protocol Editor window.
The protocol editor has the form of a table. It lists all parameters of the
protocol: names of the protocol, phase and stage, load parameters
(depending on exercise test device), intervals for the output of reports,
intervals for blood-pressure measurements).
The command menu is shown at the top of the display.
When the editor screen is opened, the cursor highlights the protocol
name (e.g. WHO). With the cursor keys, the cursor can be positioned on
the different commands.
Configuration: Stress Test
Protocol: WHO
[add phase]
del phase
print proto
add stage
del stage
reset proto
add proto
del proto
ok escape
Phase Time In Basic Lead Median Repo
STAGE
99:00
1
1. A stage time of 99.00 means an indefinite stage time and the program
does not automatically advance to the next stage
0First
PRETEST Ramping 0
00:00 0 :
EXERCISE Ramping
STAGE 1 2:00
25
2
2. Values for Ramping are 1 or Yes and 0 or No. Ramping = Yes means the
load increases evenly
1:50
STAGE 2 2:00 50 1:50
STAGE 3 2:00 75 1:50
STAGE 4 2:00 100 1:50
STAGE 5 2:00 125 1:50
STAGE 6
2:00
3
3. If the time for the first stage report is longer than the stage time, no stage
report will be printed.
150
1:50
3
STAGE 7 2:00 175 1:50
STAGE 8 2:00 200 1:50
STAGE 9 2:00 225 1:50
STAGE 10 2:00 250 1:50
STAGE 11 2:00 275 1:50
STAGE 12 2:00 300 1:50
STAGE 13 2:00 325 1:50
Protocol Editor Window