14-10 MAC 1200/1200 ST Revision C
2012250-021
Technical Specifications: Recording
Electrode Monitoring
Audible and visual indication on the LCD of disconnected electrodes or
line break; each single electrode is monitored.
Text Input
Patient and user data as well as comments can be entered via the panel
keyboard and are annotated on the recording strip.
Copy Function
In Automatic mode, after ECG recording, copies of the ECG can be
printed from memory and/or transferred to a connected PC
(configurable).
Test
Automatic performance test upon power up, including verification of the
signal path starting at the signal input.
Stored test data for demonstration of the device functions.
Power Supply
From the power line or from a built-in rechargeable battery, automatic
switchover; automatic battery charging during line-power operation from
integrated AC adapter module.
Mains Operation
Instrument design in protection class I according to IEC 60601-1
Rated voltage range 95 to 240 V
Operating voltage range 85 to 264 V, 49 to 65 Hz
Rated current 0.2 to 0.6 A
Fuse: 2 x T (slow-blow) 1.25 A, 5x20
Typical power consumption – battery charging 14 W
Maximum power consumption 29 W