Technical Specifications: Recording
Battery Operation
Type: nickel-cadmium
Rated battery voltage 18 V
Rated battery capacity 1.3 Ah
Fully charged battery sufficient for up to 50 Automatic mode 1-page
ECGs, provided recorder is only switched on to record the ECGs.
Charge time for depleted battery approximately 4 hours (minimum
charge time for 1 Automatic mode ECG: 10 minutes)
Battery life approximately 2 to 3 years, replacement by service only
Lithium battery for built-in clock, battery life approximately 5 years,
replacement by service only
Battery Maintenance
The battery in the MAC 1200/1200 ST unit should be exercised regularly.
This means the battery should be allowed to drain between charging
cycles.
If the ECG unit is left connected to wall power and/or only used
sporadically on battery power, a discharge cycle should be performed on
a monthly basis.
Turn the MAC 1200/1200 ST unit on, and allow the battery to drain.
When the Battery Low indicator illuminates, the unit can be connected
to wall power and allowed to recharge.
Use a battery maintenance chart to track the discharge cycles
Operational Readiness
After successful self-test, approximately 10 seconds after power-up.
Operating Position
Horizontal

Table of Contents

Other manuals for GE MAC 1200 ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1200 ST and is the answer not in the manual?

GE MAC 1200 ST Specifications

General IconGeneral
BrandGE
ModelMAC 1200 ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals