Section  10  —  Create,  Edit,  and  Delete  Items 
206 
To  define  or  edit  a  sensor: 
1.  Press    Options  on  the  main  menu. 
2.  Press    Sensor  Setup  >  Press  Change  Sensor  opposite  your 
chosen  measuring  channel. 
3.  Follow  EITHER  of  these  steps  according  to  your  requirements: 
•  Press    Create  New  Sensor  to  create  a  new  sensor.   
•  Use  the  arrow  keys  on  the  left  to  select  an  existing  sensor > 
Press    to  change  its  settings. 
4.  Press  the  corresponding  keys  to  define  the  settings: 
 
Toggle  the  Sensor  Type  between  acceleration, 
velocity,  displacement,  and  current. 
Toggle  the  Voltage  Ring  and  Coupling  settings. 
Toggle  Auto  Settling  ON/OFF.  When  Auto  Settling  is 
ON,  your  instrument  will  ignore  the  user-specified 
settling  time  (see  above)  and  automatically  select  the 
most  efficient  settling  time  for  your  sensor  type. 
 
Edit  the  Sensitivity.  The  range  of  sensitivities  will  vary 
depending  on  the  selected  sensor's  Store  Unit.  To 
enter  a  decimal  point,  press    then  . 
Toggle  the  Store  Unit  (e.g.  between  mV/g  and 
mV/m/s
2
  for  accelerometers). 
 
Toggle  the  sensor's  Drive  Current  ON/OFF.  Drive 
Current  is  the  2.2  mA  power  supply  for  ICP
®
-type 
accelerometers.   
 
Toggle  Bias  Voltage  Check  on  and  off  (none).  Edit, 
create,  or  delete  this  value. 
5.  Press    when  you  have  finished  editing  the  sensors.