Section  3  —  Sensors 
44 
Semi-automatic Keyphasor Threshold Selection 
 
If  a  recording  is  tach-triggered  but  no  pulses  are  being  detected,  your 
instrument  will  prompt  you  to  select  the  appropriate  Keyphasor 
threshold. 
 
If  you  are  performing  a  Tachometer  Display  measurement,  you  can  also 
bring  up  the  same  semi-automatic  selection  screen.  To  do  this:
 
 
1.  Press    or    Measure. 
2.  Select  Tachometer  Display  >  Press  .   
3.  Press    Tach  Threshold  —  Your  instrument  scans  through  the 
3V,  8V,  13V,  and  18V  threshold  levels,  displaying  the  RPM  value 
detected  at  each  one. 
4.  Use  the  keys  provided  to  select  the  threshold  that  delivers  a  stable 
RPM  reading  and  matches  the  expected  speed  of  the  machine  you 
are  testing  —  The  threshold  value  you  selected  will  be  used  for  all 
subsequent  recordings  on  the  machine  you  are  measuring. 
 
NOTE: Your instrument will increase the duration between 
thresholds to find lower RPM values. This process can take just 
over a minute. A value of zero will display if there is no RPM value.  
 
 
Correcting  the  threshold  setting  for  your  tachometer  and  machine