PEL-2000 Series User Manual
128
Set to CR Mode
Background
The PEL 2000 series load operates in three different
modes, Constant Current (CC), Constant Voltage
(CV), Constant Resistance (CR). Constant
Resistance mode will maintain a constant resistive
load, using variable current and voltage levels.
When a channel is active, the F1
key can be used to switch
between each operating mode.
Panel Operation
1. Press the CHAN button and use
the selector knob to select a
channel.
2. Press F1 until CR mode is
displayed in the display panel.
Note
Changing the operating mode will only affect the
current (active) channel. Other channels will not be
affected by any changes.
Select CR Range
Background
Constant Resistance mode can run in high and low
range. Range is dependent on the load module.
Ensure the menu is in CR Mode. See page 128.