EasyManua.ls Logo

GW Instek PEL-2004 - Recall Setup Memory (Frame Link)

GW Instek PEL-2004
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATION
229
Recall Current
Channel Preset
7. Press the Preset key and one of
the number keys.
0 = P0
1= P0 etc.
Recall All Channel
Presets
8. Press the Shift key, the Preset
key and one of the number
keys.
0 = P0
9. 1= P0 etc.
10. Press the Preset key again to
deactivate it.
PRESET
PRESET
Only the active channel’s Channel preset will be
recalled.
Recall Setup Memory (Frame link).
Background
A master mainframe can command all mainframes
(master and slave) to recall setup memory from
their internal memory. No setup data will be
recalled from the master mainframe to the slave
units.
Parameter
Setup memory 1~4.
Panel Operation
1. On the Master mainframe,
follow the procedure for
recalling setup memory for all
channels.
Page 206
All mainframes will update setup memory upon recall

Table of Contents