Configuring Step Resolution
Background
The CC, CV and CR step resolution settings can be
edited in the configuration menu. The minimum
and maximum step resolution of each channel is
dependent on the load module. For more
information on step resolution see page 81.
Parameters
Minimum* Maximum*
Unit
CCH Step HR/4000 HR/2 Amperes A
CCL Step LR/4000 LR/2 Amperes A
CRH Step HR/4000 HR/2
Siemens
CRL Step LR/4000 LR/2
Siemens
CV Step HR/4000 HR/2 Voltage V
* HR = High range, LR = Low range
Panel operation
Ensure the menu is the configuration
menu. See page 160.
1. Press the F2(Other) key to enter
the Other menu.
03/02/09
10 : 30
CC
Dyna
High
Configure
Previous
Menu
Protection
Go-NoGo
CH1
Other
S
Conf