APPENDIX
243
Battery Replacement
Background
The onboard battery is used to power the real time
clock and memory.
Step
1. Remove the screws from the rear feet (X4).
2. Remove the three screws from the side panels
(X6).
3. Pull and case back slightly and lift to remove.
4. Remove the retaining strap from the battery
socket and replace the battery.
1
2
3
4
Rating
(CR17345) Lithium manganese dioxide battery.
DL123 (Duracell Ultra 123) 1400mAh, 3.0V
(Nominal)
Disposal
Do not dispose the battery as unsorted municipal
waste. Please use a separate collection facility or
contact the supplier from which the battery was
purchased.