OPERATION
167
Adjusting the Von Voltage and Latch
Background
The Von Voltage is the voltage point at which the
load module will start to sink current. When Von
latch is set to ON, the load will continue to sink
current after being tripped, even if the voltage
drops below the Von Voltage level. The step
resolution of Von Voltage is load module
dependent.
Parameters
Von Voltage
0.0~Rating volts
Von Latch ON/OFF
Panel operation
Ensure the menu is in the
configuration menu. See page 160.
1. Press the F2(Other) key to enter
the Other menu.
03/02/09
10 : 30
CC
Dyna
High
Configure
Previous
Menu
Protection
CH1
Other Go-NoGo
Conf
2. Use the Selector knob to
highlight Von Voltage.