OPERATION
83
2. For each unit, hold the Function key while
turning the power on to enter the power on
configuration settings.
Lock/Local PROT Function Test Set
Output
Unlock ALM_CLR M 1 M 2 M 3
Shift
: Long Push
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C C
A
C V
V
VOLTAGE CURRENT
PSU 40-38
DC Power Supply
0 - 40V / 0 - 38A
Voltage Current
3. Make sure each unit is set to
Independent (F-93 = 0). When
using the power supplies in series,
each unit is operated individually,
and thus no unit is considered the
master or slave.
Page 114
Unit
F-93
Independent
0
4. Cycle the power on the units (reset the power).
Note
Configuration settings can be checked for both the
master and slave units by pressing the Function key.
Series Operation
Only operate the power supplies in series if the
units are configured correctly.
1. Turn on both units. When connected in series
unit will only show the voltage and current of
their own unit.