114 Chapter 6
P/N 130859-US, Manual Revision: AA Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
1. Close the white air line clamp and open the yellow PPP line clamp.
2. Touch Continue to begin collecting PPP.
Figure 58, Example of a Sequestration process prompt
Plasma should flow into the PPP bag at a consistent flow rate until the white
buffy coat band (made up of platelets and white cells), which is immediately
adjacent to the top of the RBC layer, reaches the shoulder of the bowl.