166 Chapter 11
P/N 130859-US, Manual Revision: AA Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Event Messages
The following section provides a list of event messages and corrective actions,
which appear on the display screen when an action is required by the user.
Note: The “Event Message Text” in the following table is displayed in software
version AQ.
Table 22, Event Messages
ID # Event Message Text
102 Effluent Line Sensor Failure
Explanation:
An issue has been detected with the effluent line sensor. The procedure can be performed, but
the device cannot monitor the effluent from the bowl.
Corrective Action:
To perform the procedure with the user monitoring the effluent:
1. Touch Continue.
2. Monitor the effluent from the bowl during processing.
If red cell spillage occurs, manually decrease the pump speed.
OR
To end the procedure:
1. Touch End Procedure.
2. If in power-on self-test, power down the device.
If the problem persists, service is required to resolve this issue.
103 Manifold Pressure Sensor Failure
Explanation:
An issue has been detected with the manifold pressure sensor.
Corrective Action:
Restart the device.
If the problem persists, service is required.

Table of Contents

Other manuals for Haemonetics Cell Saver Elite+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics Cell Saver Elite+ and is the answer not in the manual?

Haemonetics Cell Saver Elite+ Specifications

General IconGeneral
BrandHaemonetics
ModelCell Saver Elite+
CategoryMedical Equipment
LanguageEnglish

Related product manuals