166 Chapter 11
P/N 130859-US, Manual Revision: AA Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Event Messages
The following section provides a list of event messages and corrective actions,
which appear on the display screen when an action is required by the user.
Note: The “Event Message Text” in the following table is displayed in software
version AQ.
Table 22, Event Messages
ID # Event Message Text
102 Effluent Line Sensor Failure
Explanation:
An issue has been detected with the effluent line sensor. The procedure can be performed, but
the device cannot monitor the effluent from the bowl.
Corrective Action:
To perform the procedure with the user monitoring the effluent:
1. Touch Continue.
2. Monitor the effluent from the bowl during processing.
If red cell spillage occurs, manually decrease the pump speed.
OR
To end the procedure:
1. Touch End Procedure.
2. If in power-on self-test, power down the device.
If the problem persists, service is required to resolve this issue.
103 Manifold Pressure Sensor Failure
Explanation:
An issue has been detected with the manifold pressure sensor.
Corrective Action:
Restart the device.
If the problem persists, service is required.