General Operation: Sequestration 115
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual P/N 130859-US, Manual Revision: AA
Collecting PRP The device displays another set of prompts when the buffy coat starts to exit
the bowl.
1. Open the blue PRP line clamp and close the yellow PPP line clamp.
2. Touch Continue to begin collecting PRP.
Figure 59, Example of a Sequestration process prompt
The buffy coat should flow into the PRP bag until the effluent line flow from the
bowl turns medium red (indicating the presence of RBCs).