Protocol Settings 129
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual P/N 130859-US, Manual Revision: AA
Modifiable Settings
Default Settings The following tables list the Haemonetics default settings for the Cell Saver
Elite+ device:
Cell Salvage
Default Settings
Table 15, Cell Salvage Default Settings
Parameter Values Default
Notes
Fat Reduction On/Off Off
Auto-Fill On/Off On
Partial Bowl Wash Ask/Single/
Double
Ask
Pump Regulation On/Off On Pump regulation is not applicable for
the 70 mL bowl.
Quick Transfer On/Off Off This parameter is displayed only if it
has been enabled in Options. See
“Making the Quick Transfer and
Manual Mode Settings Available” on
page 136 for more information.
Manual Mode On/Off Off This parameter is displayed only if it
has been enabled in Options. See
“Making the Quick Transfer and
Manual Mode Settings Available” on
page 136 for more information.
Final Cycle: On/Off Off Only one of the sub-settings below
can be active at a time:
Last Partial Bowl Wash Double/
Single/Off
Double This sub-setting can only be
modified if Final Cycle is set to On.
End Procedure On/Off Off This sub-setting can only be
modified if Final Cycle is set to On.
Wash Line Priming On/Off On
Recentrifuge Delay On/Off On
Parameter 225 mL Bowl 125 mL Bowl 70 mL Bowl Min Max Step
Fill Start Volume (mL) 800 800 400 200 3000 100
Fill Resume Volume (mL) 400 400 200 200 3000 100