130 Chapter 7
P/N 130859-US, Manual Revision: AA Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Cell Salvage -
Manual Mode
Default Settings
Parameter 225 mL Bowl 125 mL Bowl 70 mL Bowl Min Max Step
Fill Pump Rate
a
(mL/min) 500 225 100 25 1000 25
Minimum Wash Volume (mL) 1000 750 300 500
b
5000 25
c
Wash Pump Rate
c
(mL/min) 450 200 100 25 1000 25
Smart Empty: On/Off On/Off N/A
d
N/A N/A N/A
Empty Pump Rate
e f
(mL/
min)
500/400/100 300/150/100 100
d
25 1000 25
Emergency Empty Pump
Rate
e f
(mL/min)
500/400/250 300/250/100 N/A
d
25 1000 25
Emergency Fill Pump Rate
(mL/min)
800 400 N/A 25 1000 25
Emergency Wash Pump Rate
(mL/min)
800 400 N/A 25 1000 25
a. If pump regulation is on, the device optimizes the quality of the final product by adjusting the pump
speed during the Fill phase from 150 - 500 mL/min (150-225 mL/min for a 125 mL bowl), during the Wash
phase from 150-450 mL/min (150-200 mL/min for a 125 mL bowl), during the Concentrate phase from 125-
150 mL/min (75-125 mL/min for a 125 mL bowl) and during a Wash phase after a Concentrate phase from
100-200 mL/min (75-175 mL/min for a 125 mL bowl) based upon the effluent line sensor readings.
b. Minimum wash volume for the 125 mL bowl is 375 mL; for the 70 mL bowl it is 210 mL.
c. Step is 10 mL for the 70 mL bowl.
d. For the 70 mL bowl, the Smart Empty and Emergency Empty Pump Rate settings are not available, but
the Empty Pump Rate setting is always available.
e. If Smart Empty is on, the device reduces the pump speed during the Empty/Return phase in
preprogrammed increments to minimize the number of RBCs remaining in the bowl at the end of the
Empty/Return phase.
f. For the 225 mL and 125 mL bowls, Empty Pump Rate and Emergency Empty Pump Rate are only
available for modification when Smart Empty is On, otherwise, they are grayed out.
Table 15, Cell Salvage Default Settings
Table 16, Cell Salvage - Manual Mode Default Settings
Parameter Value Default
Recentrifuge Delay On/Off On
Alert Sounds On/Off On