Protocol Settings 131
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual P/N 130859-US, Manual Revision: AA
Sequestration
Default Settings
Pump Rate (mL/min) 225 mL
Bowl
125 mL
Bowl
70 mL
Bowl
Min Max Step
Fill Pump Rate 500 500 125 25 1000 25
Wash Pump Rate 500 500 100 25 1000 25
Empty Pump Rate 500 500 100 25 1000 25
Conc Pump Rate 150 150 75 25 1000 25
Return Pump Rate 500 500 100 25 1000 25
Centrifuge Speed (RPM) 225 mL
Bowl
125 mL
Bowl
70 mL
Bowl
Min Max Step
Fill Centrifuge Speed 5650 5650 7500 2050 (2100 for
the 70 mL bowl)
5650 (7500 for
the 70 mL bowl)
100
Wash Centrifuge Speed 6050 6050 7500 2050 (2100 for
the 70 mL bowl)
6050 (7500 for
the 70 mL bowl)
100
Conc Centrifuge Speed 3850 3850 5000 2050 (2100 for
the 70 mL bowl)
5650 (7500 for
the 70 mL bowl)
100
Table 16, Cell Salvage - Manual Mode Default Settings
Table 17, Sequestration Default Settings
Parameter Values Default
PPP Return Plasma No/Yes No
PRP Milk No/Yes No
Parameter 225 mL Bowl 125 mL Bowl 70 mL Bowl Min Max Step
Fill Pump Rate (mL/min) 60 60 N/A 10 250 10
Fill Centrifuge Speed (RPM) 5650 5650 N/A 2050 5650 50
PRP Collection Pump Rate
(mL/min)
20 20 N/A 10 250 10
PRP Collection Centrifuge
Speed (RPM)
2450 2450 N/A 2050 5650 50
Empty Pump Rate
f
(mL/min)
500/400/100 300/150/100 N/A 25 1000 25
PRP Extend Volume (mL) 10 10 N/A 0 50 1
PRP Milk Volume (mL) 3 3 N/A 1 50 1
PRP Milk Delay (seconds) 10 10 N/A 1 30 1

Table of Contents