136 Chapter 7
P/N 130859-US, Manual Revision: AA Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Making the
Quick Transfer
and Manual
Mode Settings
Available
Two additional modes can be made available on the Cell Saver Elite+ device:
Quick Transfer Mode enables the device to quickly move unwashed shed
blood from the reservoir to the reinfusion bag and bypass the wash phase
completely when time is of the essence.
Manual Mode enables the device to be operated manually by the user.
Note: These modes are only available when using the Cell Salvage protocol.
Note: These modes can only be made available in a settings group. To create
a new settings group, see “Creating a New Settings Group” on page 125.
Note: Contact your local Haemonetics representative for access to training and
the password to make this feature available
To make the Quick Transfer or Manual Mode setting available:
1. Touch (Menu).
2. Select System from the drop-down list. The System screen appears.
3. Touch Unlock. The Enter Access Code screen appears.
4. Enter the access code and touch (Accept).
5. On the System screen, touch Options. The Option screen appears.
6. Touch either Quick Transfer or Manual Mode.
7. Touch (Up) or (Down) to change the setting to Yes. If desired,
select the other mode and change its setting to Yes.
8. Touch Done.
The enabled functionality is now available for use in the settings group. To
utilize Quick Transfer, see “Emergency Options” on page 85 and “Quick
Transfer Mode” on page 86. To utilize Manual Mode, see “Manual Mode” on
page 88.