Troubleshooting 165
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual P/N 130859-US, Manual Revision: AA
Touch Screen
Problems
Device Cover
Problems
Table 20, Troubleshooting
Problem Possible Cause Corrective Action
The touch pads are
unresponsive or your
touches are detected
inaccurately.
The screen needs
to be recalibrated.
1. While in the Ready screen,
press the (Stop) key
rapidly until the Calibration
screen appears.
2. Follow the prompts to calibrate
the touch screen and touch
(Accept) when complete.
3. Restart the device.
Table 21, Troubleshooting
Problem Possible Cause Corrective Action
The device cover
does not unlock.
On rare
occasions, the
device cover
does not unlock
when expected
after an event
message
appears.
1. Press the (Stop) key
located on the device display.
The device cover unlocks.
2. Resolve any issue reported by
the event message.