EasyManua.ls Logo

Haemonetics Cell Saver Elite+ - Electromagnetic Compatibility

Haemonetics Cell Saver Elite+
230 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IEC/EN 60601-1-2 Standard Requirements 217
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual P/N 130859-US, Manual Revision: AA
Electromagnetic Compatibility
Note: There were no deviations from IEC/EN 60601-1-2 or allowances used
during testing of the system.
Note: The emissions characteristics of this equipment make it suitable for use
in industrial areas and hospitals (CISPR11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this equipment
might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating
or re-orienting the equipment.
Table 23, Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The system is intended for use in the electromagnetic environment specified below. The customer or
user of the system should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment Guidance
RF emissions
CISPR 11
Group 1 The system uses RF energy only for its
internal functions. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The system is suitable for use in all
establishments, other than domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC/EN 61000-3-3
Complies

Table of Contents

Other manuals for Haemonetics Cell Saver Elite+

Related product manuals