134 Chapter 7
P/N 130859-US, Manual Revision: AA Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Wash Line Priming (On/Off)
Determines if the device omits the Wash Prime process at the beginning of the
first Fill phase to reduce the duration of the procedure.
On: The device performs the Wash Prime process at the beginning of the
first Fill phase.
Off: The device performs the Wash Prime process at the beginning of the
first Wash phase.
Recentrifuge Delay (On/Off)
Determines if device performs the recentrifuge delay. For more information,
see “Recentrifuge Delay” on page 91.
Note: This must be set separately for Cell Salvage and Manual Mode.
On: The device performs the recentrifuge delay. The user can override
the recentrifuge delay after 15 seconds by touching (Play).
Off: The device does not perform the recentrifuge delay (the device does
not wait for the centrifuge to ramp up before starting the pump).
Attention: If the delay is turned off, be aware of the possibility of red cell
spillage caused by pumping fluid into a bowl of red blood cells with poor
or no separation.
Parameters Cell Salvage
Fill Start Volume: The approximate volume of fluid in the reservoir that
transitions the device from Standby to Fill when the bowl is empty.
Fill Resume Volume: The approximate volume of fluid in the reservoir
that transitions the device from Standby to Fill when the bowl already
contains some fluid.
Fill Pump Rate: The approximate rate at which the pump turns while
filling the bowl with fluid.
Minimum Wash Volume: The minimum volume of wash solution to be
used to wash one bowl of fluid.
Wash Pump Rate: The approximate rate at which wash solution enters
the bowl.
Smart Empty: Determines if the device slows the pump speed at a
decremental rate to empty fluid from the bowl or if the pump speed
remains constant. This setting is not available for the 70 mL bowl.
On: The pump empties fluid from the bowl at a preprogrammed
decremental speed determined by the parameter settings.
Off: The pump empties fluid at a constant speed, which the user can
adjust via the +/- pads from 25 mL/min to 1000 mL/min (default is 100
mL/min).