14 Chapter 1
P/N 130859-US, Manual Revision: AA Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Cell Saver Elite+
Release AQ - Key
New Features
Table 1, Key New Features
See the Cell Saver Elite+ Release Notes for more information on the changes
made for release AQ.
Performance
Characteristics
Feature Description
Manual Mode
Manual Mode enables users to operate the
device manually during the Fill, Wash, Empty,
Concentrate and Return phases.
Quick Transfer When time is of the essence, Quick Transfer
Mode can be used to quickly move shed blood
from the reservoir to the reinfusion bag and
bypass the wash phase completely.
Final Cycle
Enables users to specify the End Procedure
behavior when air is detected during Fill.
Table 2, Vacuum Suction
Characteristic Value
Smart Suction
Operating vacuum 30 to 165 mmHg ±10%
(4.0 to 22.0 kPa; 40.0 to 220 mbar)
Vacuum cutoff 175 mmHg
(23.3 kPa; 233 mbar)
Maximum free air flow 40 L/min
Manual Mode
Operating vacuum 50 to 250 mmHg
(6.7 to 33.3 kPa; 66.7 to 333.3 mbar)
Maximum free air flow 40 L/min
Post-Op
Operating vacuum
25 to100
(3.3 to 13.3 kPa; 33.3 to 133.3 mbar)
Maximum free air flow 40 L/min